Sample: |
Whole blood, serum, plasma |
Measurement Wavelength: |
5 wavelengths (405, 550, 575, 610, and 820 nm) |
Measurement Principle: |
Optical measurement of reflection intensity of reagent color reaction |
Processing Speed: |
48 Samples per hour |
Minimum Sample Volume: |
250 µL (whole blood), 100 µL (serum) |
Simultaneous Measurement: |
3 Single Reagent Strips and/or 1 Multi Reagent Strip |
Reagent Reaction Temperature: |
37° C |
Light Source: |
LED and Interference filter |
Calibration Method: |
Calibration by magnetic card |
Data Storage Volume: |
100 tests |
Display: |
20 digits x 2 lines |
Built-in Printer: |
36 Character thermal printer (58mm width) |
External Output: |
RS-232C interface |
Power Supply: |
AC100~120V, 220~240V, 50/60Hx (AC adapter method) |
Power Input: |
Max. 100VA |
Dimensions: |
Width 13.5" x Depth 8" x Height 6.5" |
Weight: |
12 lbs. |